view our radiation-shielding
solutions

On this page you will find our product range for radiopharmaceutical packaging. We display these products well structured, clearly visualized and complemented with product sheets. It will give you a comprehensive insight in the radiation-shielding solutions we offer. 

Modular hot cell

  • Master slave manipulation system
  • Large lead glass window
  • Highly configurable options
  • High speed access door
  • Large interior space

The modular hot cell model MHC is a very versatile hot cell. The MHC can be configured to meet any laboratory layout with various hot cell connections, large lead glass window providing full view of the interior, air environmental control, multiple product manipulation and product retrieval options. The MHC is precisely engineered and constructed for continuous protection and reliable everyday laboratory use. The design of the MHC allows for the use of various master slave manipulators.  
 
As a standard this cell is fitted with shielding for the ionization chamber and basic gas connections. Behind the vertical motor driven lead shielded door, the hinged acrylic door maintains airtight integrity. The main compartment is shielded with 75 mm of lead on all sides. The internal stainless steel box with rounded corners is easy to clean.  
 

Several options such as various gas taps, laptop support and ion chamber lift are available for the MHC. 

Model VG-MHC

  • High speed access door
  • Large interior space
  • Master slave manipulation system
  • Integrated ion chamber shielding
  • Highly configurable options

To optimize the configuration of the modular hot cell to your specific needs, the following options are available: 

  • Master-slave manipulators, CRL type G-LDR
  • Lead plugs instead of master-slave manipulators
  • Extended gas regulation
  • Storage cabinet beneath main compartment
  • Solid and or liquid waste vault
  • Digital pressure gauge instead of analog gauge
  • Internal HEPA/Charcoal filter
  • External shielded HEPA/Charcoal filter
  • Ionization chamber shield for dose calibrator, 50 mm / 2 inch or 75 mm / 3 inch lead
  • Ionization chamber lift
  • Foot pedal control for ionization chamber lift
  • Universal support (outlet, cable pass through, arm for laptop or ion chamber control unit)
  • Glove ports in shielded front door with hinged plugs
  • Glove ports in acrylic door
  • Radiation detection system including safety interlock on front door
  • Extended cover plates to specified height